Generalized low-rank nonrigid motion-corrected reconstruction for MR fingerprinting

dc.contributor.authorCruz, Gastao
dc.contributor.authorQi, Haikun
dc.contributor.authorJaubert, Olivier
dc.contributor.authorKuestner, Thomas
dc.contributor.authorSchneider, Torben
dc.contributor.authorMichael Botnar, Rene
dc.contributor.authorPrieto, Claudia
dc.date.accessioned2024-01-10T13:45:02Z
dc.date.available2024-01-10T13:45:02Z
dc.date.issued2021
dc.description.abstractPurpose: Develop a novel low-rank motion-corrected (LRMC) reconstruction for nonrigid motion-corrected MR fingerprinting (MRF).
dc.description.abstractMethods: Generalized motion-corrected (MC) reconstructions have been developed for steady-state imaging. Here we extend this framework to enable nonrigid MC for transient imaging applications with varying contrast, such as MRF. This is achieved by integrating low-rank dictionary-based compression into the generalized MC model to reconstruct MC singular images, reducing motion artifacts in the resulting parametric maps. The proposed LRMC reconstruction was applied for cardiac motion correction in 2D myocardial MRF (T-1 and T-2 ) with extended cardiac acquisition window (similar to 450 ms) and for respiratory MC in free-breathing 3D myocardial and 3D liver MRF. Experiments were performed in phantom and 22 healthy subjects. The proposed approach was compared with reference spin echo (phantom) and with 2D electrocardiogram-triggered/breath-hold MOLLI and T-2 gradient-and-spin echo conventional maps (in vivo 2D and 3D myocardial MRF).
dc.description.abstractResults: Phantom results were in general agreement with reference spin-echo measurements, presenting relative errors of approximately 5.4% and 5.5% for T-1 and short T-2 (<100 ms). respectively. The proposed LRMC MRF reduced residual blurring artifacts with respect to no MC for cardiac or respiratory motion in all cases (2D and 3D myocardial, 3D abdominal). In 2D myocardial MRF, left-ventricle T-1 values were 1150 +/- 41 ms for LRMC MRF and 1010 +/- 56 ms for MOLLI; T-2 values were 43.8 +/- 2.3 ms for LRMC MRF and 49.5 +/- 4.5 ms for T-2 gradient and spin echo. Corresponding measurements for 3D myocardial MRF were 1085 +/- 30 ms and 1062 +/- 29 ms for T-1, and 43.5 +/- 1.9 ms and 51.7 +/- 1.7 ms for T-2. For 3D liver, LRMC MRF measured liver T-1 at 565 +/- 44 ms and liver T-2 at 35.4 + 2.4 ms.
dc.description.abstractConclusion: The proposed LRMC reconstruction enabled generalized (nonrigid) MC for 2D and 3D MRF, both for cardiac and respiratory motion. The proposed approach reduced motion artifacts in the MRF maps with respect to no motion compensation and achieved good agreement with reference measurements.
dc.description.funderEngineering and Physical Sciences Research Council (EPSRC)
dc.description.funderWellcome EPSRC Center for Medical Engineering
dc.description.funderDepartment of Health through the National Institute for Health Research comprehensive Biomedical Research Center award award to Guy's St Thomas' National Health Service (NHS) Foundation Trust in partnership with King's College London and King's College H
dc.fechaingreso.objetodigital2024-04-09
dc.format.extent18 páginas
dc.fuente.origenWOS
dc.identifier.doi10.1002/mrm.29027
dc.identifier.eissn1522-2594
dc.identifier.issn0740-3194
dc.identifier.pubmedidMEDLINE:34601737
dc.identifier.urihttps://doi.org/10.1002/mrm.29027
dc.identifier.urihttps://repositorio.uc.cl/handle/11534/78974
dc.identifier.wosidWOS:000702807500001
dc.information.autorucFacultad de Ingeniería; Prieto Vasquez, Claudia Del Carmen; S/I; 14195
dc.language.isoen
dc.nota.accesocontenido completo
dc.publisherWILEY
dc.revistaMAGNETIC RESONANCE IN MEDICINE
dc.rightsacceso abierto
dc.subject2D cardiac
dc.subject3D liver
dc.subject3D cardiac
dc.subjectlow rank
dc.subjectMR fingerprinting
dc.subjectnonrigid motion correction
dc.subjectMYOCARDIAL T-1
dc.subjectSENSE
dc.subject.ods03 Good Health and Well-being
dc.subject.odspa03 Salud y bienestar
dc.titleGeneralized low-rank nonrigid motion-corrected reconstruction for MR fingerprinting
dc.typeartículo
sipa.codpersvinculados14195
sipa.indexWOS
sipa.trazabilidadCarga SIPA;09-01-2024