We report a 26 year old patient who had a single chamber pacemaker implantation one year before. During a routine pre-operative evaluation, pacemaker dysfunction was demonstrated due to sensing and pacing failure, associated to left pectoral muscle rhythmic contraction. Chest X-ray confirmed Twiddler syndrome, in which twisting or rotation of the device inside the pocket results in lead dislodgement and device malfunction. (Rev Med Chile 2011; 139: 917-919).
Registro Sencillo
Registro Completo
Autor | Ramirez E, Pablo Barrero, Raul, V Gonzalez F, Rodrigo Frangini S, Patricia Vergara S, Ismael |
Título | Twiddler syndrome. Report of one case |
Revista | REVISTA MEDICA DE CHILE |
ISSN | 0034-9887 |
Volumen | 139 |
Número de publicación | 7 |
Página inicio | 917 |
Página final | 919 |
Fecha de publicación | 2011 |
Resumen | We report a 26 year old patient who had a single chamber pacemaker implantation one year before. During a routine pre-operative evaluation, pacemaker dysfunction was demonstrated due to sensing and pacing failure, associated to left pectoral muscle rhythmic contraction. Chest X-ray confirmed Twiddler syndrome, in which twisting or rotation of the device inside the pocket results in lead dislodgement and device malfunction. (Rev Med Chile 2011; 139: 917-919). |
Derechos | acceso abierto |
DOI | 10.4067/S0034-98872011000700014 |
Editorial | SOC MEDICA SANTIAGO |
Enlace | |
Id de publicación en Pubmed | MEDLINE:22051831 |
Id de publicación en WoS | WOS:000293874000014 |
Paginación | 3 páginas |
Palabra clave | Arrhythmias, cardiac Pacemaker, artificial Twiddler syndrome |
Tema ODS | 03 Good Health and Well-being |
Tema ODS español | 03 Salud y bienestar |
Tipo de documento | artículo |